Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Mapping point-of-care performance using locally-smoothed median and maximum absolute difference curves
Gerald J Kost1, Nam K Tran, Harpreet Singh
1Point-of-Care Testing Center for Teaching and Research, Pathology and Laboratory Medicine, School of Medicine, University of California, Davis, CA 95616, USA. gjkost@ucdavis.edu
Background:
The goal is to introduce visual performance mapping efficient for establishing acceptance criteria and facilitating decisions regarding the utility of hospital point-of-care devices. This approach uniquely reveals the quality of performance locally, as opposed to globally.
Methods:
After presenting theoretical foundations, this study illustrates the approach by applying it to six hospital glucose meter systems (GMSs) using clinical multi-center (n=2767) and multi-system (n=613, n=100) observations.
Results:
LS MAD curves identified breakouts, that is, points where the locally-smoothed median absolute difference (LS MAD) curve exceeds the recommended error tolerance limit of 5 mg/dL (0.28 mmol/L). LS maximum absolute difference (MaxAD) breakthroughs, which occur where the LS MaxAD curve exceeds the 99th percentile of MaxADs from x=30-200 mg/dL (1.67-11.10 mmol/L), showed extreme error locations. A multi-sensor interference- and hematocrirt-correcting GMS displayed a flat LS MAD curve until it reached a breakout of 179 mg/dL (9.94 mmol/L) and generated breakthroughs that could affect bedside decision-making, but less erratically than other systems with inadequate performance for hospital critical care. We discovered Class I (meter high, reference low) and Class II (converse) discrepant values in some systems. Class I errors could lead to inappropriate insulin dosing and hypoglycemic episodes in tight glucose control.
Conclusions:
LS MAD-MaxAD curves help assess the performance of point-of-care testing. Visual mapping of systematic and random errors locally over the entire analyte measurement range in a single integrated display is an advantage when considering the adverse impact of zones of poor quantitative performance on specific clinical applications, threshold-driven bedside decisions and the care of critically ill patients.
Related Concept Videos
Mean Absolute Deviation
Let us consider a dataset containing the number of unsold cupcakes in five shops: 10, 15, 8, 7, and 10. Initially, calculate the sample mean. Then calculate the deviation, or the difference, between each data value and the mean. Next, the absolute values of these deviations are added and divided by the sample size to...
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...
Midpoint Rule
Central Tendency: Analysis
The mean is one such measure, calculated by totaling all values in a dataset and dividing by the number of values. For instance, the mean blood pressure reading (120, 130, 140, 150) would be 135. However, the mean can be affected by extreme values or outliers.
The median, another measure,...
Receiver Operating Characteristic Plot
Residuals and Least-Squares Property
If the observed data point lies above the line, the residual is positive, and the line underestimates the actual data value for y. If the observed data point lies below the line, the residual is negative, and the line overestimates the actual data value for y.
The process of fitting the best-fit...