Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Evaluation of the implanted cochlear implant electrode by CT scanning with three-dimensional reconstruction
Wei-Jia Kong1, Hua-Mao Cheng, Hui Ma
1Department of Otorhinolaryngology, Union Hospital of Tongji Medical College, Huazhong University of Science and Technology, 1277 Jiefang Avenue, Hankou, Wuhan, Hubei, PR China. entwjkong@yahoo.com.cn
Conclusions:
Computed tomography (CT) scan with three-dimensional (3D) reconstruction of the inner ear provides a more accurate image of the relationship of the electrode within the cochlear canal, with direct demonstration of electrode insertion depth in the cochlea in comparison with X-ray plain film.
Objective:
This study was designed to evaluate the value of spiral CT scans with 3D reconstruction in determining the insertion site and depth of implanted cochlear implant electrodes.
Methods:
A total of 172 cochlear implant recipients were involved in this study. The implanted electrodes of all patients were examined by X-ray plain film, and 157 cochlear recipients were examined by spiral CT scans with axial 1 mm image slices. The data from the CT scans were transferred to a workstation for 3D reconstruction (direct volume rendering) of the inner ear. The pseudocolor technique was used to display the electrode.
Results:
The insertion depth of the electrode could be evaluated indirectly by the X-ray plain film. In contrast, the stereoscopic images from a CT scan with 3D reconstruction of the inner ear demonstrated the shape, position, and insertion depth of the electrode more accurately.