Related Experiment Video
Updated: May 27, 2026

Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
Results of the first randomized French study evaluating self-testing of the International Normalized Ratio
Kasra Azarnoush1, Lionel Camilleri, Bruno Aublet-Cuvelier
1Heart Surgery Department, Hôpital Gabriel Montpied CHU Clermont-Ferrand, France. kazarnoush@chu-Clermontferrand.fr
Background And Aim Of The Study:
In this single-center French study, the conventional management of patients receiving vitamin K antagonists (VKAs) was compared with an International Normalized Ratio (INR) self-testing program. The aim was to determine the reliability of self-testing devices, and to estimate the variability of the self-measured INR within the therapeutic and target range.
Methods:
A total of 206 patients who had undergone valve replacement with a mechanical prosthesis, with or without myocardial revascularization, between May 2004 and September 2007 was randomized into two groups. Group 1 patients (n = 103) underwent INR monitoring at a laboratory, while Group 2 patients (n = 103) underwent self-testing INR using either the CoaguChek (Roche) (Group 2A; n = 55) or INRatio (Hemosense) (Group 2B; n = 48) system. Patients in Group 1 underwent at least once-monthly INR measurement, while those in Group 2 carried out once-weekly self-testing, and also underwent once-monthly INR measurement at the laboratory. The large majority of patients (97.9%) were treated with fluindione.
Results:
The mean follow up period was 49.0 +/- 10.3 weeks. Self-testing was reliable, with a correlation coefficient between device- and laboratory-measured INRs of 0.80 [CI: 0.78, 0.82] (p < 0.0001). The proportion of time spent within the INR target range was significantly higher for Group 2 (61.5 +/- 19.3% versus 55.5 +/- 19.9%; p < 0.05), while the absolute mean deviation of INR from the target range was higher in Group 1 (60.1 +/- 70.2% versus 47.4 +/- 51.5%). Adverse events were reported by seven patients in Group 1, but by no patients in Group 2 (p < 0.01).
Conclusion:
INR self-testing devices are reliable and beneficial. Moreover, INR self-monitoring allows an enhanced stability within a target range, and also helps to prevent serious postoperative complications.