Aim:
To identify the optimal sensor application method that gave the quickest display of accurate heart rate (HR) data using the Nellcor OxiMax N-600x pulse oximeter (PO).
Methods:
Stable infants who were monitored with an electrocardiograph were included. Three sensor application techniques were studied: (i) sensor connected to cable, then applied to infant; (ii) sensor connected to cable, applied to investigator's finger, and then to infant; (iii) sensor applied to infant, then connected to cable. The order of techniques tested was randomized for each infant. Time taken to apply the PO sensor, to display data and to display accurate data (HR(PO) = HR(ECG) ± 3 bpm) were recorded using a stopwatch.
Results:
Forty infants were studied [mean (SD) birthweight, 1455 (872) g; gestational age, 31 (4) weeks; post-menstrual age, 34 (4) weeks]. Method 3 acquired any data significantly faster than methods 1 (p = 0.013; CI, -9.6 to -3.0 sec) and 2 (p = 0.004; CI, -5.9 to -1.2 sec). Method 3 acquired accurate data significantly faster than method 1 (p = 0.016; CI, -9.4 to -1.0 sec), but not method 2 (p = 0.28).
Conclusion:
Applying the sensor to the infant before connecting it to the cable yields the fastest acquisition of accurate HR data from the Nellcor PO.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...