A high-resolution 3D ultrasonic system for rapid evaluation of the anterior and posterior segment
Gholam A Peyman1, Charles P Ingram, Leonardo G Montilla
1Department of Radiology, University of Arizona, Tucson, Arizona, USA. gpeyman1@yahoo.com
Background And Objective:
Traditional ultrasound imaging systems for ophthalmology employ slow, mechanical scanning of a single-element ultrasound transducer. The goal was to demonstrate rapid examination of the anterior and posterior segment with a three-dimensional (3D) commercial ultrasound system incorporating high-resolution linear probe arrays.
Materials And Methods:
The 3D images of the porcine eye were generated in approximately 10 seconds by scanning one of two commercial linear arrays (25- and 50-MHz). Healthy enucleated pig eyes were compared with those with induced injury or placement of a foreign material (eg, metal). Rapid, volumetric imaging was also demonstrated in one human eye in vivo.
Results:
The 50-MHz probe provided exquisite volumetric images of the anterior segment at a depth up to 15 mm and axial resolution of 30 μm. The 25-MHz probe provided a larger field of view (lateral X depth: 20 × 30 mm), sufficient for capturing the entire anterior and posterior segments of the pig eye, at a resolution of 60 μm. A 50-MHz scan through the human eyelid illustrated detailed structures of the Meibomian glands, cilia, cornea, and anterior segment back to the posterior capsule.
Conclusion:
The 3D system with its high-frequency ultrasound arrays, fast data acquisition, and volume rendering capability shows promise for investigating anterior and posterior structures of the eye.