Is P(br)O (2) pressure reactivity index (ORx) dependent on the type of oxygen probe? An in vivo study
G Grözinger1, Martin Schenk, Christian Thiel
1Department of General and Transplantation Surgery, Eberhard-Karls-University Hospital, Tübingen, Germany.
Objective:
To evaluate if ORx is dependent on the type of brain tissue O(2) (P(br)O(2)) probe in an in vivo setting.
Methods:
In eight German landrace pigs two types of probes were implanted simultaneously in the same cerebral hemisphere. All pigs underwent hepatectomy and received neuromonitoring until death. A LICOX(®) probe CCI.S, representing a Clarke type electrode, was compared with a Raumedic Neurovent PTO, representing an optode. Data were sampled at 50 Hz. Average values were calculated every 30 s. Cerebral perfusion pressure (CPP) was averaged over 30 s. ORx was calculated for each probe. To increase the signal to noise ratio of the ORx, the ORx values, which had been assessed every minute, were averaged over 1 h.
Results:
The overall measurement time was 145.1 h (8,703 data pairs). Despite a mean difference of 6.2 mmHg (p < 0.0001) in the measured values of P(br)O(2), the mean ORx(licox) was 0.139, mean ORx(raumedic) 0.146 (p = 0.2098). Correlation coefficient of ORx values assessed every minute and every hour was 0.52 and 0.58 respectively.
Conclusion:
Despite this significant difference in absolute values of P(br)O(2) the derived mean ORx values were not different. Similar to the established Licox system, the Raumedic system seems to enable a valid ORx recording.
Related Concept Videos
Oxygen Transport in the Blood
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated. Under...
Regioselectivity of Electrophilic Additions-Peroxide Effect
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.