An Ex Vivo Porcine Model for Hydrodynamic Testing of Experimental Aortic Valve Procedures and Novel Medical Devices
Published on: August 25, 2023
In vivo testing of an intra-annular aortic valve annuloplasty ring in a chronic calf model
J Scott Rankin1, Jeff L Conger, Egemen Tuzun
1Centennial Medical Center, Vanderbilt University, Nashville, TN 37205, USA. jsrankinmd@cs.com
Objective:
To increase applicability and stability of aortic valve repair, a three-dimensional aortic annuloplasty ring has been developed for intra-annular placement. The goal of this study was to test the safety of this device with in vivo implantation in the calf model.
Methods:
In 10 chronic calves, the HAART annuloplasty ring was sutured to the aortic valve annulus using cardiopulmonary bypass. The animals were recovered and followed for 1-2 months. Serial echocardiography was used to evaluate valve competence, and contrast aortograms and CT angiograms were obtained in selected animals. After completion of follow-up, each animal was euthanized, and aortic endoscopy was performed under water distension in five. Full autopsies with histologic examinations were performed.
Results:
All animals survived surgery. Two were euthanized in the first week for complications, and the remaining eight calves were followed uneventfully for the 1-2 months. Serial echocardiography showed completely competent valves in all but one animal, in which the ring was intentionally up-sized to test the sizing strategy. Contrast aortographic and CT angiographic findings were similar to the echocardiograms. Postmortem examination showed proper seating of all rings with endothelialization at 1-2 months. All valves demonstrated good leaflet coaptation and no abnormalities.
Conclusions:
In vivo testing of a three-dimensional aortic annuloplasty ring in a chronic calf model proved to be very successful and safe. Using the sizing and implant strategies developed, human trials seem indicated.