Development of high-voltage pulse-slicer unit with variable pulse duration for pulse radiolysis system
J Upadhyay1, M L Sharma, C P Navathe
1Laser Electronic Support Division, Raja Ramanna Centre for Advanced Technology, Indore, Madhya Pradesh 452013, India. janky@rract.gov.in
Abstract:
A high-voltage pulse-slicer unit with variable pulse duration has been developed and integrated with a 7 MeV linear electron accelerator (LINAC) for pulse radiolysis investigation. The pulse-slicer unit provides switching voltage from 1 kV to 10 kV with rise time better than 5 ns. Two MOSFET based 10 kV switches were configured in differential mode to get variable duration pulses. The high-voltage pulse has been applied to the deflecting plates of the LINAC for slicing of electron beam of 2 μs duration. The duration of the electron beam has been varied from 30 ns to 2 μs with the optimized pulse amplitude of 7 kV to get corresponding radiation doses from 6 Gy to 167 Gy.