Robotic Cochlear Implantation for Direct Cochlear Access
Published on: June 16, 2022
A self-developed and constructed robot for minimally invasive cochlear implantation
Brett Bell1, Christof Stieger, Nicolas Gerber
1ARTORG Center, University of Bern, Switzerland.
Conclusion:
A robot built specifically for stereotactic cochlear implantation provides equal or better accuracy levels together with a better integration into a clinical environment, when compared with existing approaches based on industrial robots.
Objectives:
To evaluate the technical accuracy of a robotic system developed specifically for lateral skull base surgery in an experimental set-up reflecting the intended clinical application. The invasiveness of cochlear electrode implantation procedures may be reduced by replacing the traditional mastoidectomy with a small tunnel slightly larger in diameter than the electrode itself.
Methods:
The end-to-end accuracy of the robot system and associated image-guided procedure was evaluated on 15 temporal bones of whole head cadaver specimens. The main components of the procedure were as follows: reference screw placement, cone beam CT scan, computer-aided planning, pair-point matching of the surgical plan, robotic drilling of the direct access tunnel, and postoperative cone beam CT scan for accuracy assessment.
Results:
The mean accuracy at the target point (round window) was 0.56 ± 0.41 mm with an angular misalignment of 0.88 ± 0.40°. The procedural time for the registration process through the completion of the drilling procedure was 25 ± 11 min. The robot was fully operational in a clinical environment.