Study of arteriovenous fistula function by the temperature gradient method using the Twister™ device
Ramon Roca-Tey1, Rosa Samon, Omar Ibrik
1Hospital de Mollet, Mollet del Vallès, Barcelona, Spain.
Summary
The temperature gradient method (TGM) with the Twister™ device accurately measures arteriovenous fistula (AVF) blood flow in hemodialysis (HD) patients. This method is faster than the Delta-H technique and shows no correlation with patient blood pressure.
Area of Science:
- Nephrology
- Vascular Surgery
- Medical Devices
Background:
- Periodic blood flow (QA) measurement is crucial for arteriovenous fistula (AVF) surveillance in chronic hemodialysis (HD) patients.
- Assessing AVF function helps prevent complications and ensure effective treatment.
Purpose of the Study:
- To evaluate the efficacy of the temperature gradient method (TGM) using the Twister™ device for determining AVF blood flow (QA).
- To compare the TGM's functional results with the established Delta-H method.
- To investigate the influence of blood pressure on AVF function.
Main Methods:
- Non-invasive QA measurement in 30 AVF patients during HD using TGM with the Twister™ device.
- QA calculation via blood temperature monitor (BTM) readings at normal and reverse blood line configurations.
- Simultaneous measurement of mean arterial pressure (MAP).
- Comparison with AVF blood flow determined by the Delta-H method.
Main Results:
- The TGM with Twister™ demonstrated reproducible QA measurements (CV 6.8 +/- 4.7%).
- Mean QA was 1132.5 +/- 515.4 ml/min, with no significant difference between diabetic and non-diabetic patients.
- No correlation was found between QA and age, time on HD, MAP, Kt/V, needle distance, or AVF duration.
- TGM with Twister™ (5.9 min) was significantly faster than Delta-H (26.8 min) (p < .001).
- TGM-derived QA highly correlated with Delta-H measurements (r = 0.963, p < .001).
Conclusions:
- The TGM is a valuable, reproducible method for QA measurement during HD.
- The Twister™ device significantly reduces the time required for TGM QA measurements.
- TGM and Delta-H methods yield highly correlated AVF blood flow values.
- AVF function, as measured by blood flow, is not significantly related to patient blood pressure.
Related Concept Videos
Hemodialysis I: Introduction
Hemodialysis (HD) is a medical treatment that artificially removes waste products, excess fluids, and toxins from the blood when the kidneys are no longer able to perform these functions effectively. In this process, blood is filtered through a semipermeable membrane, allowing for the selective removal of waste while preserving necessary components like blood cells and proteins. Hemodialysis is typically performed in patients with end-stage renal disease (ESRD) or severe kidney...
Assessing Body Temperature - Temporal Artery
Here is a stepwise guide to assessing the body temperature at the temporal artery using a temporal artery thermometer
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...