Related Experiment Video
Updated: May 23, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Multicenter experience with transvenous lead extraction of active fixation coronary sinus leads
Melanie Maytin1, Roger G Carrillo, Pablo Baltodano
1Cardiovascular Division, Brigham and Women's Hospital, Boston, Massachusetts 02115, USA. mmaytin@partners.org
Background/Objective:
Active fixation coronary sinus (CS) leads limit dislodgement and represent an attractive option to the implanter. Although extraction of passive fixation CS leads is a common and frequently uncomplicated procedure, data regarding extraction of chronically implanted active fixation CS leads are limited.
Methods:
We performed a retrospective cohort study of patients undergoing active fixation CS lead extraction at six centers. Patient and procedural characteristics, indications for extraction, use of extraction sheath (ES) assistance, and outcomes are reported.
Results:
Between January 2009 and February 2011, 12 patients underwent transvenous lead extraction (TLE) of Medtronic StarFix® lead (Medtronic Inc., Minneapolis, MN, USA). The cohort was 83% male with mean age 71 ± 14 years. Average implant duration was 14.2 ± 5.7 months (2.3-23.6). All leads but one were removed for infectious indications (67% systemic infection). At the time of explant, the fixation lobes were completely retracted in only one of the 12 cases and ES assistance was required for lead removal in all cases (58% laser, 25% cutting, 25% mechanical, and 25% femoral). The majority of cases required advancement of the sheath into the CS (75.0%) and often into a branch vessel (41.7%). One lead could not be removed transvenously and required surgical lead extraction. There were no major complications. Examination of the leads after extraction frequently revealed significant tissue growth into the fixation lobes.
Conclusions:
Although TLE of active fixation CS leads can be a safe procedure in select patients and experienced hands, powered sheaths and aggressive techniques are frequently required for successful removal despite relatively short implant durations. This raises significant concern regarding future TLE of active fixation CS leads with longer implant durations.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization