Anterior High-Resolution Optical Coherence Tomography in the Diagnosis and Therapeutic Monitoring of Ocular Surface Squamous Neoplasia
Published on: August 9, 2024
Corneal thickness mapping by 3D swept-source anterior segment optical coherence tomography
Alberto Neri1, Marianna Malori, Patrizia Scaroni
1Ophthalmology, University Hospital of Parma, Parma, Italy. neri.mail2@gmail.com
Swept-source anterior segment optical coherence tomography (AS-OCT) offers excellent repeatability for central corneal thickness (CCT) measurements and pachymetric mapping. While AS-OCT, spectral-domain retinal OCT, and ultrasound pachymetry show high correlation, AS-OCT provides superior repeatability.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Medical Imaging
Background:
- Accurate central corneal thickness (CCT) measurement is crucial for diagnosing and managing various ocular conditions.
- Different pachymetry devices, including optical coherence tomography (OCT) and ultrasound, are used for CCT assessment.
- Evaluating the accuracy and repeatability of these devices is essential for clinical practice.
Purpose of the Study:
- To compare the accuracy and repeatability of central corneal thickness (CCT) measurements using swept-source anterior segment optical coherence tomography (AS-OCT), spectral-domain retinal OCT with a corneal module, and ultrasound pachymetry (USP).
- To assess the repeatability of pachymetric mapping provided by AS-OCT.
Main Methods:
- Fifty healthy volunteers underwent repeated CCT measurements using AS-OCT, retinal OCT, and USP by a single operator.
- Statistical analyses included repeated-measures ANOVA, Bonferroni tests, Pearson correlation, and Bland-Altman plots.
- Repeatability was evaluated using Cronbach's Alpha, intraclass correlation coefficient (ICC), and coefficient of repeatability.
Main Results:
- Mean CCT values were 540±28.9 μm (AS-OCT), 544±29.5 μm (retinal OCT), and 549.3±31.7 μm (USP), with statistically significant differences (p<0.01).
- All three instruments demonstrated high correlation in CCT measurements (r > 0.96).
- AS-OCT exhibited significantly higher repeatability for CCT measurements (p<0.001) and excellent repeatability for pachymetric maps (ICC=0.999).
Conclusions:
- Swept-source AS-OCT provides highly repeatable central corneal thickness (CCT) measurements and excellent repeatability for pachymetric mapping.
- While AS-OCT, retinal OCT, and USP measurements are highly correlated, they are not identical.
- AS-OCT demonstrates superior repeatability compared to the other assessed methods.