Evaluation of the stability of iron(II) solutions by precise coulometric titration with electrogenerated cerium(IV)
Toshiaki Asakai1, Akiharu Hioki
1National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Tsukuba, Ibaraki, Japan. t-asakai@aist.go.jp
Abstract:
An iron(II) solution is often used as a reducing agent in titrimetry and standardized with cerium(IV) or potassium dichromate. Such an iron(II) standard solution is needed for not only titrimetric analyses, but also instrumental ones. Iron(II) is unstable even in a highly acidic solution, mainly due to air-oxidation; therefore, its standardization is required before use. In the present study, the concentration of an iron(II) solution was accurately determined by coulometric titration with electrogenerated cerium(IV), and also by gravimetric titration with a standard potassium dichromate; new useful information concerning the stability of iron(II) solutions in aqueous sulfuric acid was obtained. The current efficiency of the coulometric titration with electrogenerated cerium(IV) was not very high; however, it was found that the titration efficiency was sufficient to assay an iron(II) solution.
Related Concept Videos
Controlled-Current Coulometry: Coulometric Titration
The fundamental requirements for coulometric titrations are (1) 100% efficiency in the reagent-generating electrode reaction and (2) a stoichiometric and preferably rapid reaction between the generated...
Complexometric Titration: Overview
Controlled-Current Coulometry: Overview
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides and...
Complexometric EDTA Titration Curves