Malfunction of the Siaretron 1000 Iper™ hyperbaric ventilator
Pieter A Bothma1, Andreas Brodbeck, Uday Bapat
1Consultant in Anaesthesia and Intensive Care and lead consultant of the East of England Hyperbaric Unit, James Paget University Hospitals NHS Foundation Trust, Lowestoft Road, Gorleston, Norfolk NR31 6LA, United Kingdom, Phone: +44-(0)1493-452475, Fax: +44-(0)1493-452753,
Introduction:
Patient ventilators for use in the hyperbaric chamber need to be of special design; any malfunction could have disastrous consequences. We report two serious problems with a recently purchased Siaretron 1000 Iper™ ventilator.
Methods:
The ventilator was tested with a Biotek VT-Plus™ gas flow analyzer, which also measures O2 concentration. The changes in fraction of inspiratory oxygen (FiO2) were verified with a Teledyne Electronic Devices™ O2 analyzer.
Results:
In volume control ventilation (VCV) mode: excessively large tidal volumes were delivered when the fraction of inspiratory oxygen (FiO2) was changed. In pressure control ventilation (PCV) mode: changing the FiO2 setting did not change the FiO2 delivered by the ventilator. The ventilator also exhibited an irregular flow pattern in PCV.
Conclusions:
These problems may cause serious diagnostic and clinical consequences if not identified as equipment malfunction issues. A malfunction of the integrated memory in the microchip on the main board was said to cause the PCV malfunction. The manufacturer replaced the main board, which corrected the problem. The solution offered for the VCV problem was to change FiO2 in steps of 0.1 per breath, which eliminates the tidal volume surges. We feel it is extremely important that all users of the Siaretron 1000 Iper™ are made aware of these problems as they are not described in the user manual or elsewhere.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Acute Respiratory Failure-III
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...