Characterization and Application of Passive Samplers for Monitoring of Pesticides in Water
Published on: August 3, 2016
Pneumatic nebulization gas-solid extraction of some pesticides in liquor
Yeqiang Wang1, Fan Wang, Cui Yu
1College of Chemistry, Jilin University, 2699 Qianjin Street, Changchun 130012, China.
Abstract:
In general, solid phase extraction (SPE) was applied to the separation and concentration of analytes from liquid samples. However, the SPE generally cannot be applied directly to the separation and concentration of analytes from liquor in which the concentration of ethanol is high, because the ethanol can elute the analytes from the SPE column and the analytes can hardly been adsorbed onto the column. In the work, the pneumatic nebulization coupled with gas-solid extraction (PN-GSE) was applied to the extraction of several pesticides, including fenuron, chlortoluron, karmex, linuron and prebane, from liquor samples. The high concentration of ethanol in the liquor samples cannot elute the analytes from the SPE column. The high performance liquid chromatography was applied to the separation and determination of the analytes. Experimental parameters affecting the performances of PN-GSE, such as types and amount of packed material of the column, temperature of the spray chamber, flow rate of carrier gas, pumping rate of gas and volume of elution solvent were examined and optimized. The limits of detection for the analytes were in the range of 0.19-0.26 μg L⁻¹. Liquor samples were analyzed and the recoveries of the analytes in spiked samples were from 91.0 to 106.3%. The relative standard deviations were lower than 6.5%.