Sample Preparation for Probe Electrospray Ionization Mass Spectrometry
Published on: February 19, 2020
Electrospray ionisation mass spectrometric studies for the determination of palladium after pre-concentration by
P G Jaison1, Pranaw Kumar, Vijay M Telmore
1Fuel Chemistry Division, Bhabha Atomic Research Centre, Trombay, Mumbai-400 085, India.
Rationale:
Electrospray ionisation mass spectrometric (ESI-MS) analysis of Pd in complex matrices is difficult due to the multiplicity of matrix effects. Two different approaches, internal standard and matrix separation, were investigated for developing a reliable analytical procedure for the trace level determination of Pd in simulated high-level liquid waste (SHLLW) solutions.
Methods:
An ESI mass spectrometer with a quadrupole-time-of-flight analyser was used to study the speciation of the palladium-benzoylthiourea (Pd-BTU) complex and to determine the Pd content. The Pd-BTU complex was selectively pre-concentrated using disposable pipette extraction (DPX). Extraction parameters as well as ESI-MS parameters such as concentration of BTU, acidity, composition of medium and capillary voltage, etc., were optimized based on the major species [Pd(BTU)(2)S](+).
Results:
The method gave quantitative and selective pre-concentration of the Pd-BTU complex from SHLLW. Linearity from 5 ppb to 200 ppb and a limit of detection of 0.012 ppb were obtained for Pd. No interference from the neighboring elements, viz. ruthenium, rhodium, silver and cadmium, was observed during the determination of Pd based on the [Pd(BTU)(2)S](+) peak. The ESI signal intensity was not influenced by the presence of the many other elements in the SHLLW solution.
Conclusions:
Good sensitivity, tolerance to matrix concentration and the absence of interference from neighboring elements make the method very promising for the determination of Pd at low levels in complex samples. We have demonstrated the capability of ESI-MS for the quantification of Pd in complex matrices and its potential for providing data on speciation, using the Pd-BTU complex.
More Related Videos
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
10:44Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview