Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Artifact reduction from metallic dental materials in T1-weighted spin-echo imaging at 3.0 tesla
Sang-Young Zho1, Min-Oh Kim, Keun-Woo Lee
1Department of Electrical and Electronic Engineering, Yonsei University, Seoul, Korea.
Purpose:
To investigate and propose a method of artifact reduction arising from metallic dental materials by applying a slice-encoding for metal artifact correction (SEMAC) technique on T1-weighted spin-echo (SE) imaging at 3 Tesla.
Materials And Methods:
The view angle tilting (VAT) technique was adapted to conventional T1-weighted spin-echo (SE) sequence to correct the in-plane distortion, and the SEMAC technique was used for correcting the remaining through-plane distortions. Fourier transform based B0 field simulations were performed to estimate the amount of field perturbation and a scout imaging method was developed which guide in selecting the number of slice-encodings needed in SEMAC sequences. Phantoms of six different dental materials with various shapes and sizes that are used in practice (amalgam; titanium implant; gold and Ni-Cr crowns; Ni-Ti and stainless steel orthodontic wires) were imaged. In vivo images of two subjects were also acquired. The amounts of artifact reduction were quantified in phantom studies.
Results:
Compared with conventional SE imaging in phantom studies, in-plane artifacts were reduced by up to 43% in the VAT SE images and 80% in the SEMAC images. Through-plane artifacts were reduced by up to 65% in SEMAC images. In vivo SEMAC images also showed reduced artifacts.
Conclusion:
The SEMAC technique can mitigate artifact caused by metallic dental materials for T1w-SE imaging.