Discrepancies in determining electron energy for lumpectomy boost treatment
1St Vincent Hospital Green Bay WI, USA.
Abstract:
The aim of this study was to compare lumpectomy cavity depth measurements obtained through ultrasound (U/S) and retrospective computed tomography (CT). Twenty-five patients with stage T1-2 invasive breast cancer formed the cohort of this study. Their U/S and CT measurements were converted into electron energy and compared. The mean U/S depth was 3.6 ± 1.3 cm, while the mean CT depth was 4.9 ± 1.9 cm; the listed error ranges are one standard deviation. Electron energies for treatment ranged from 6 MeV to 12 MeV based on the U/S determination. There was no significant correlation between cavity depths measured by U/S and CT (R(2)= 0.459, P < 0.002). Furthermore, only 20% of CT-based electron energy determinations matched the corresponding U/S determinations. This ratio increased to 40% when taking into account an upper limit based on the depth of organs at risk below the cavity. The study shows that there is a significant discrepancy between cavity depths determined by U/S and CT. It also supports the concept that post-lumpectomy radiotherapy boosts should be tailored according to the needs and comfort of individual practices and institutions.