Related Experiment Video
Updated: May 17, 2026

A Coupled Experiment-finite Element Modeling Methodology for Assessing High Strain Rate Mechanical Response of Soft Biomaterials
Published on: May 18, 2015
Chinks in the armor: percutaneous injuries from hollow bore safety-engineered sharps devices
1University of Nevada, Reno, Reno, NV, USA. lblack@unr.edu
Background:
Despite their overwhelming efficacy, safety-engineered sharp devices (SESDs) cause a residual fraction of injuries. Although the fraction of injuries from SESDs is less than that reported for nonsafety devices, it remains a "preventable fraction" and is a sizable target for further advances.
Methods:
A retrospective review of 3,297 percutaneous injuries from hollow bore safety-engineered devices occurring between 2001 and 2009 was conducted examining the Exposure Prevention Information Network (EPINet) needlestick surveillance data.
Results:
Nurses sustain 64.6% of all SESD injuries. 42.9% Of SESD injuries occur after device use and are likely preventable through consistent and effective use of safety-engineered technology. Excluding injuries that occurred during device use or between procedural steps, 71.8% (n/N = 28/39) of physician injuries, 58.2% (n/N = 645/1,109) of injuries to nurses, and 45.8% (n/N = 88/192) of injuries to phlebotomists occurred when an available SESD was not fully activated.
Conclusion:
Passive devices that do not require action on the part of the end user to engage a safety feature currently represent a small portion of the SESD market. Wider dissemination of a broader array of passive SESDs coupled with continual education of end users is essential to an effective sharps injury prevention program.
Related Concept Videos
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus.
Endocarditis I: Introduction
Cardiac Catheterization I: Pre-Procedure Overview
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...