Split calibration curve: an approach to avoid repeat analysis of the samples exceeding ULOQ
Sudipta Basu1, Abdul Basit, Selvan Ravindran
1Department of DMPK & Toxicology, Sai Advantium Pharma Ltd, Building 1, Plot 2, Chrysalis Enclave, International Biotech Park, Phase 2, Hinjewadi, Pune, 411057, India. basu.sudipta@gmail.com
Background:
The current practice of using calibration curves with narrow concentration ranges during bioanalysis of new chemical entities has some limitations and is time consuming. In the present study we describe a split calibration curve approach, where sample dilution and repeat analysis can be avoided without compromising the quality and integrity of the data obtained.
Results:
A split calibration curve approach is employed to determine the drug concentration in plasma samples with accuracy and precision over a wide dynamic range of approximately 0.6 to 15,000 ng/ml for dapsone and approximately 1 to 25,000 ng/ml for cyclophosphamide and glipizide. A wide dynamic range of concentrations for these three compounds was used in the current study to construct split calibration curves and was successfully validated for sample analysis in a single run.
Conclusion:
Using this method, repeat analysis of samples can be avoided. This is useful for the bioanalysis of toxicokinetic studies with wide dose ranges and studies where the sample volume is limited.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Calibration Curves: Correlation Coefficient
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Blank Solutions
In some experimental cases, the reagents, solvents, or lab equipment used in...