Related Experiment Video
Updated: May 17, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Extending the detection window of diazepam by directly analyzing its glucuronide metabolites in human urine using
Xin Wang1, Rong Wang, Yurong Zhang
1Shanghai Institute of Pharmaceutical Industry, State Key Laboratory of New Drug and Pharmaceutical Process, Shanghai 200040, China.
Abstract:
A method for the simultaneous direct analysis of diazepam oxazepam glucuronide, temazepam glucuronide, oxazepam, nordiazepam, and temazepam in human urine was developed and validated. Urine sample was purified by solid phase extraction (SPE), and the analysis was achieved using a liquid chromatography-tandem mass spectrometry (LC-MS/MS) system equipped with an electrospray ionization source (ESI). Multiple reaction monitoring (MRM) mode was used to analyze the target compounds. Extraction recoveries were 65-122% for all the analytes. The method showed acceptable intra-assay and inter-assay precision (both relative standard deviation (RSD)≤11.2%) for quality control (QC) samples. The limits of detections (LODs) were in the range of 0.1-2 ng/mL. The present assay was applied to analyze the urine obtained from three volunteers after oral administration of a single dose 5mg of diazepam. The results showed that, the detection periods of oxazepam glucuronide and temazepam glucuronide were much longer than diazepam and other metabolites.