Performance of independent dose calculation in helical tomotherapy: implementation of the MCSIM code
Eric Dzintars1, Sotirios Stathakis, Panayiotis Mavroidis
1Department of Radiological Sciences, University of Texas Health Science Center, San Antonio, TX, USA.
Abstract:
Currently, a software-based second check dose calculation for helical tomotherapy (HT) is not available. The goal of this study is to evaluate the dose calculation accuracy of the in-house software using EGS4/MCSIM Monte Carlo environment against the treatment planning system calculations. In-house software was used to convert HT treatment plan information into a non-helical format. The MCSIM dose calculation code was evaluated by comparing point dose calculations and dose profiles against those from the HT treatment plan. Fifteen patients, representing five treatment sites, were used in this comparison. Point dose calculations between the HT treatment planning system and the EGS4/MCSIM Monte Carlo environment had percent difference values below 5 % for the majority of this study. Vertical and horizontal planar profiles also had percent difference values below 5 % for the majority of this study. Down sampling was seen to improve speed without much loss of accuracy. EGS4/MCSIM Monte Carlo environment showed good agreement with point dose measurements, compared to the HT treatment plans. Vertical and horizontal profiles also showed good agreement. Significant time saving may be obtained by down-sampling beam projections. The dose calculation accuracy of the in-house software using the MCSIM code against the treatment planning system calculations was evaluated. By comparing point doses and dose profiles, the EGS4/MCSIM Monte Carlo environment was seen to provide an accurate independent dose calculation.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses