Automated T(2) * measurements using supplementary field mapping to assess cardiac iron content
Brian A Taylor1, Ralf B Loeffler, Ruitian Song
1Department of Radiological Sciences, St. Jude Children's Research Hospital, Memphis, TN 38105, USA.
Purpose:
To develop and evaluate an algorithm that automatically identifies high-susceptibility areas and excludes them from T(2) * measurements in the left ventricle (LV) for myocardial iron measurements.
Materials And Methods:
An autoregressive moving average (ARMA) model was implemented on multigradient echo scans of 24 patients (age range 3-45 years, 10 male/14 female). Voxels with relatively high susceptibility (>3 Hz/mm) were flagged and deselected from the T(2) * calculations for iron quantification. The mean, standard deviation, and coefficient of variation (CoV) of the ARMA-defined region were compared to the CoV of four distinct regions of the LV and the entire LV using a Student's t-test (α = 0.05).
Results:
The CoV of T(2) * values obtained by the ARMA method are comparable with that in the interventricular septum (IS), where susceptibility was the lowest (CoV = 0.31). The ARMA method provides a greater area (51.9 ± 13.7% of the LV) to measure T(2) * than that using the IS alone (21.1 ± 3.4%, P < 0.0001). Areas where low susceptibility are measured corroborate with areas reported in previous studies that investigated T(2) * variations throughout the LV.
Conclusion:
An automated method to measure T(2) * relaxation in the LV with minimal effects from susceptibility has been developed. Variability is reduced while covering more regions for cardiac T2 * calculation.