Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT--phantom study
Touko Kaasalainen1, Kirsi Palmu, Anniina Lampinen
1HUS Medical Imaging Center, Helsinki University Central Hospital, POB 340, Haartmaninkatu 4, 00290, Helsinki, Finland. touko.kaasalainen@hus.fi
Background:
CT optimization has a special importance in children. Smaller body size accentuates the importance of patient positioning affecting both radiation dose and image quality.
Objective:
To determine the effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT examination.
Materials And Methods:
Chest scans of a pediatric 5-year anthropomorphic phantom were performed in different vertical positions (-6 cm to +5.4 cm) with a 64-slice CT scanner. Organ doses were measured with metal-oxide-semiconductor field-effect transistor (MOSFET) dosimeters. Image noise and contrast were determined from the CT number histograms corresponding to different tissues.
Results:
Significant changes in organ doses resulting from vertical positioning were observed, especially in radiosensitive anterior organs. The breast dose increased up to 16% and the thyroid dose up to 24% in lower positions. The noise was increased up to 45% relative to the centre position in the highest and lowest vertical positions, with a particular increase observed on the anterior and posterior sides, respectively. Off-centering also affected measured image contrast.
Conclusion:
Vertical off-centering markedly affects organ doses and measured image-quality parameters in pediatric chest CT examination. Special attention should be given to correct patient centering when preparing patients for CT scans, especially when imaging children.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography