Short communication: the effect of temperature on performance of milk ketone test strips
J Shire1, J L Gordon, E L Karcher
1Department of Animal Science, Michigan State University, East Lansing, MI 48824, USA.
Abstract:
Ketosis is estimated to affect 15% of early lactation dairy cows. A ketone test strip (Keto-Test; Elanco Animal Health, Greenfield, IN) allows producers a method to determine the concentration of β-hydroxybutyrate (BHBA) in milk to track individual animal and herd incidence of ketosis. The objective of this study was to determine the effect of altering the temperature of milk and the test strips at the time of the test on the reliability of the Keto-Test. A total of 118 Holstein cows, ranging from 5 to 17 DIM, were selected from a commercial Holstein dairy herd in Michigan. A milk sample was collected from the right rear quarter of each cow during the a.m. milking. Each sample was tested under 4 temperature conditions: (1) Keto-Test strips and milk at room temperature (RT; 24.0 ± 0.1°C; control; manufacturer's instructions), (2) cold strips (10.8 ± 0.9°C) and milk at RT, (3) cold strips and fresh milk, and (4) strips at RT and fresh milk. Milk was recorded as negative (0-99 μmol/L), weak positive (100-199 μmol/L), positive (200-499 μmol/L), or highly positive (≥ 500 μmol/L). Blood samples were collected immediately following milk collection and analyzed for BHBA concentration using a ketone test meter. Cows with blood BHBA concentration of ≥ 1,400 μmol/L were considered positive for subclinical ketosis. Accuracy of the Keto-Test strips under the 4 conditions was determined by the κ coefficient of agreement, using the result of condition 1 as the accepted true value. Additionally, sensitivity and specificity were calculated using the blood BHBA concentrations and results of each of the 4 conditions. Using the Keto-Test 60.2% of cows tested negative for milk BHBA, 24.6% tested weak positive, 14.4% tested positive, and 0.8% tested highly positive. The weighted κ coefficient of agreement between the control condition (1) and condition 2, 3, and 4 and 95% lower and upper confidence intervals were as follows: condition 2=0.71 (0.62, 0.80), condition 3=0.69 (0.60, 0.78), and condition 4=0.63 (0.54, 0.73). These results indicate good agreement between the outcome of condition 1 and conditions 2, 3, and 4. The sensitivities/specificities for 1, 2, 3, and 4 were as follows: 0.77/0.79, 0.74/0.75, 0.69/0.88, and 0.69/0.84, indicating that the test in all temperature conditions had a strong ability to detect the presence of BHBA in milk. In conclusion, the reliability of the Keto-Test strips was not dependent on the temperature of the milk or the test strips.
Related Concept Videos
Effect of Temperature Change on Reaction Rate
Effects of Temperature on Free Energy
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Factors Affecting Body Temperature
Factors may include:
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...