Parametric study of antennas for long range Doppler radar heart rate detection
Mehran Baboli1, Aditya Singh, Noah Hafner
1Electrical Engineering Department of University of Hawaii at Manoa, Honolulu, HI 96822, USA. Baboli@hawaii.edu
Abstract:
This research presents results obtained from long range measurements of physiological motion pertaining to human cardiac and respiration activity. A pulse pressure sensor was used as reference to verify the results from radar signals. A motion detection and grading algorithm was used to detect the presence of heart rate. In addition to showing that human heart rate and respiration can be measured at distances of 21 and 69 meters respectively, the effect of antenna size, radiation pattern and gain on the range of the radar has also been studied.