Related Experiment Videos
Accuracy and precision of methods for theophylline measurement in physicians' offices
J D Cook1, G E Platoff, T R Koch
1Department of Pathology, University of Maryland School of Medicine, Baltimore 21201.
Abstract:
We tested the accuracy and precision of five theophylline methods intended for use in physicians' offices. The Syntex AccuLevel, Ames Seralyzer, and 3M Diagnostics TheoFAST methods were less reproducible (CVs 6.3% to 9.2%) than the Abbott Vision and Kodak DT-60 (CVs 2.2% to 3.3%). Caffeine interfered with the Vision, Seralyzer, and AccuLevel methods, and theobromine interfered with the Vision, Seralyzer, and TheoFAST methods. Only the DT-60 method was free from interference from any of the 24 compounds tested. Results by all methods correlated well with those by the HPLC comparison method (Clin Chem 1981;27:1931-3) and by the Abbott TDx method for assay of 100 serum (or, when appropriate, paired whole-blood) samples. The frequency of sample results differing from the comparison method by greater than 2.0 mg/L was as follows: TDx, 11%; Vision (serum), 12%; Vision (whole blood), 18%; DT-60, 14%; AccuLevel, 18%; Seralyzer, 25%; and TheoFAST, 31%. The Kodak DT-60 method was the most nearly accurate and precise among these physician's office methods. Some physician's office methods for theophylline analysis are not adequate to guide dosage adjustments.