Spinal Sonography for Ultrasound-Guided Lumbar Neuraxial Anesthesia
Published on: January 31, 2025
A pilot study to compare epidural identification and catheterization using a saline-filled syringe versus a
Yasser M Samhan1, Hossam H El-Sabae, Hanan F Khafagy
1Department of Anesthesia, Theodor Bilharz Research Institute, Ministry of Scientific Research, P.O. Box 30, Warak El-Hadar, Kornish El-Nile, Imbaba, Giza, 12411, Egypt.
Journal of Anesthesia
|February 15, 2013
Summary
A new continuous hydrostatic pressure system significantly reduces the time and attempts needed to identify the epidural space. This novel technique offers an easier method for epidural anesthesia procedures in adults.
Area of Science:
- Anesthesiology
- Neurosurgery
- Medical Devices
Background:
- Epidural anesthesia is a common procedure for surgical anesthesia and analgesia.
- Accurate identification of the epididural space is crucial for successful and safe epidural block placement.
- Conventional methods, such as the loss of resistance technique with a saline-filled syringe, can be time-consuming and require multiple attempts.
Purpose of the Study:
- To introduce and evaluate a novel continuous hydrostatic pressure system for epidural space identification and catheterization.
- To compare the efficacy and safety of the new pressure technique against the conventional saline-filled syringe technique.
Main Methods:
- A prospective randomized study involving 108 adult patients undergoing elective endoscopic urological procedures.
- Patients were randomized to either the conventional saline-filled syringe (group C) or the new pressure technique (group P) for loss of resistance epidural technique.
- The pressure technique involved observing free flow of pressurized normal saline (50 mmHg) through the epidural needle during advancement.
Main Results:
- The pressure technique group (P) showed a significant reduction in the time to identify the epidural space (20s vs. 60.5s, p=0.001) and the number of attempts (1 vs. 1, p=0.02) compared to the conventional group (C).
- A significant reduction in motor block was also observed in group P (1 vs. 2, p=0.02).
- No significant differences were found in epidural catheterization ease, sensory block, anesthesia quality, or side effects between the groups.
Conclusions:
- The new continuous hydrostatic pressure system provides an easier and more efficient method for identifying the epidural space.