Related Experiment Video
Updated: May 11, 2026

09:01
Treatment of Ankle Osteoarthritis with Total Ankle Replacement Through a Lateral Transfibular Approach
Published on: January 24, 2018
Comparative study of medial opening-wedge high tibial osteotomy using 2 different implants
Woon-Hwa Jung1, Chung-Woo Chun, Ji-Hoon Lee
1Department of Orthopedic Surgery, Murup Hospital, Gyeongnam, South Korea.
Summary
The Aescula short spacer plate for opening-wedge high tibial osteotomy (HTO) showed a higher complication rate and loss of correction compared to the TomoFix plate over 2 years. The TomoFix plate demonstrated superior outcomes.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Sports medicine
Background:
- Opening-wedge high tibial osteotomy (HTO) is a surgical procedure to correct knee alignment.
- The choice of fixation plate can influence surgical outcomes and complication rates.
- Evaluating novel fixation devices is crucial for improving patient results.
Purpose of the Study:
- To compare the clinical and radiographic outcomes of opening-wedge HTO using a short spacer plate (Aescula) versus a rigid long plate (TomoFix).
- To assess complication rates, including plate-related issues and loss of correction, at 2-year follow-up.
Main Methods:
- Retrospective analysis of 94 opening-wedge HTOs with the Aescula plate (Group I) and 92 with the TomoFix plate (Group II).
- Clinical evaluation using Knee Society scores and radiographic assessment of mechanical tibiofemoral angle (mTFA) preoperatively and at 2-year follow-up.
- Complication rates were systematically recorded and compared between the two groups.
Main Results:
- Group II (TomoFix) had a significantly lower overall complication rate (26%) and plate-related complication rate (9%) compared to Group I (Aescula) (38% and 20%, respectively).
- While both groups showed improvement in Knee Society scores, Group II achieved significantly higher postoperative scores.
- Group II also demonstrated better maintenance of the mechanical tibiofemoral angle at follow-up.
Conclusions:
- The Aescula short spacer plate is associated with a higher incidence of plate-related complications and loss of correction after opening-wedge HTO compared to the TomoFix plate.
- The rigid TomoFix plate appears to provide more stable fixation and better long-term correction in this patient cohort.
- Further investigation into the biomechanical properties and long-term efficacy of short spacer plates in HTO is warranted.