Immediate loading with LASER-treated surface implant: two-year resonance frequency analysis follow-up
Marco Duvina1, Luigi Barbato, Jacopo Buti
1Department and Postgraduate of Oral Surgery, Florence University Dental School, Florence, Italy.
The Journal of Craniofacial Surgery
|May 30, 2013
Summary
LASER-treated dental implants show increasing stability over 24 months in patients with complete upper jaw edentulism. Resonance frequency analysis (RFA) confirms sustained implant stability quotient (ISQ) values throughout the study.
Area of Science:
- Dental Implantology
- Biomaterials Science
- Oral Surgery
Background:
- Implant surface characteristics critically influence osteointegration and primary stability.
- LASER surface treatment offers precise control over roughness, ensuring a repeatable and uncontaminated implant surface.
- Optimized surface properties are essential for successful osseointegration and long-term implant function.
Purpose of the Study:
- To evaluate the long-term implant stability of LASER-treated surfaces using repeated resonance frequency analysis (RFA).
- To assess implant behavior over 24 months in patients with complete upper jaw edentulism under an immediate-loading protocol.
- To correlate RFA measurements with osseointegration and stability following immediate loading.
Main Methods:
- Ten patients received 6-8 LASER-surface implants each, with torque values monitored during insertion.
- All implants were immediately loaded within 24 hours with a temporary full-arch prosthesis.
- Resonance frequency analysis (RFA) and radiographic controls (OPT) were performed at baseline and at 1, 3, 6, 12, and 24 months post-loading.
Main Results:
- Implant stability quotient (ISQ) values showed a statistically significant decrease from baseline to 1 month (P = 0.0039).
- A significant increase in ISQ was observed from 1 to 3 months (P = 0.0156), 3 to 6 months (P = 0.0020), 6 to 12 months (P = 0.020), and 12 to 24 months (P = 0.0391).
- Rapid ISQ increments were noted in early bone remodeling phases, indicating successful osseointegration.
Conclusions:
- LASER-treated surface implants demonstrated favorable implant stability quotient (ISQ) values throughout the 24-month follow-up.
- The findings align with existing literature on RFA analysis in immediate loading protocols for the upper jaw.
- RFA is a reliable method for monitoring implant stability in immediate loading scenarios.