Related Experiment Video
Updated: May 10, 2026

16:52
Test Samples for Optimizing STORM Super-Resolution Microscopy
Published on: September 6, 2013
Simple buffers for 3D STORM microscopy
Nicolas Olivier1, Debora Keller, Vinoth Sundar Rajan
1Laboratory for Experimental Biophysics, School of Basic Sciences, Swiss Federal Institute of Technology (EPFL), CH-1015 Lausanne, Switzerland.
Biomedical Optics Express
|June 14, 2013
Summary
Researchers found that Vectashield, a common mounting medium, simplifies 3D STORM super-resolution microscopy. This method achieves high resolution without complex buffers, improving reproducibility and accessibility for advanced imaging techniques.
Area of Science:
- Biophysics
- Optical Microscopy
- Nanotechnology
Background:
- 3D STORM (3D Super-resolution Orientation-Resolved Microscopy) is a powerful super-resolution imaging technique.
- Achieving high resolution (10 nm lateral, 30-50 nm axial) requires precise control over dye molecule blinking.
- Current methods often rely on complex, sensitive buffers and enzymatic systems, hindering reproducibility and accessibility.
Purpose of the Study:
- To investigate the suitability of a commercial mounting medium for 3D STORM imaging.
- To determine if a simpler buffer system can yield comparable or superior results to complex, traditional buffers.
- To reduce barriers to entry and improve the reproducibility of 3D STORM experiments.
Main Methods:
- Utilized the commercial mounting medium Vectashield for 3D STORM imaging of Alexa-647 dye molecules.
- Compared imaging performance and resolution against established complex buffer systems.
- Focused on evaluating image quality, especially in the axial (3D) dimension.
Main Results:
- Vectashield enabled successful 3D STORM imaging of Alexa-647.
- Images obtained using Vectashield were comparable or superior to those from complex buffers.
- The commercial mounting medium demonstrated particular effectiveness for 3D imaging applications.
Conclusions:
- Vectashield serves as an effective and simpler alternative buffer for 3D STORM imaging.
- This finding lowers the barrier to entry for 3D STORM, promoting wider adoption.
- The use of Vectashield offers a more reproducible method for comparing optical setups and data processing algorithms in super-resolution microscopy.