[FE modeling and analysis of the locking plate for proximal humerus fracture]
Feng Lin1, Tianping Yao, Chengtao Wang
1Shanghai Inspection and Testing Institute for Medical Devices, Shanghai, 201321.
Abstract:
The FE model of proximal humeral locking plate was established and analyzed to provide academic support of the proximal humeral locking plate failure. The stress distributions of the locking plate were observed under the situations of the fracture area providing support and without support. In summary, this type of locking plate cannot afford the stress at 120 degrees abduction position under the situation of fracture area provides without support. In order to reduce the probability of locking plate failure, it is recommended that using other type of locking plate instead or avoiding 120 degree abduction of the patient.
Related Concept Videos
Bones of the Upper Limb: Humerus
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a short...
Bones of the Upper Limb: Ulna
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...