Ion profiling in an ambient drift tube-ion mobility spectrometer using a high pixel density linear array detector
Stephen J Davila1, Omar Hadjar, Gary A Eiceman
1Department Chemistry and Biochemistry, New Mexico State University, Las Cruces, New Mexico 88003, United States. sjdavila@nmsu.edu
Abstract:
A linear pixel-based detector array, the IonCCD, is characterized for use under ambient conditions with thermal (<1 eV) positive ions derived from purified air and a 10 mCi (63)Ni foil. The IonCCD combined with a drift tube-ion mobility spectrometer permitted the direct detection of gas phase ions at atmospheric pressure and confirmed a limit of detection of 3000 ions/pixel/frame established previously in both the keV (1-2 keV) and the hyper-thermal (10-40 eV) regimes. Results demonstrate the "broad-band" application of the IonCCD over 10(5) orders in ion energy and over 10(10) in operating pressure. The Faraday detector of a drift tube for an ion mobility spectrometer was replaced with the IonCCD providing images of ion profiles over the cross-section of the drift tube. Patterns in the ion profiles were developed in the drift tube cross-section by control of electric fields between wires of Bradbury Nielson and Tyndall Powell shutter designs at distances of 1-8 cm from the detector. Results showed that ion beams formed in wire sets, retained their shape with limited mixing by diffusion and Coulombic repulsion. Beam broadening determined as 95 μm/cm for hydrated protons in air with moisture of ~10 ppmv. These findings suggest a value of the IonCCD in further studies of ion motion and diffusion of thermalized ions, enhancing computational results from simulation programs, and in the design or operation of ion mobility spectrometers.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
07:103D Depth Profile Reconstruction of Segregated Impurities Using Secondary Ion Mass Spectrometry
Published on: April 29, 2020
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Capillary Electrophoresis: Instrumentation
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
High-Performance Liquid Chromatography: Types of Detectors
MALDI-TOF Mass Spectrometry