Related Experiment Video
Updated: May 8, 2026

Electrically Conductive Scaffold to Modulate and Deliver Stem Cells
Published on: April 13, 2018
Electric field stimulation through a biodegradable polypyrrole-co-polycaprolactone substrate enhances neural cell
Hieu T Nguyen1, Shawn Sapp, Claudia Wei
1Department of Biomedical Engineering, The University of Texas at Austin, Austin, Texas, 78712.
Abstract:
Nerve guidance conduits (NGCs) are FDA-approved devices used to bridge gaps across severed nerve cables and help direct axons sprouting from the proximal end toward the distal stump. In this article, we present the development of a novel electrically conductive, biodegradable NGC made from a polypyrrole-block-polycaprolactone (PPy-PCL) copolymer material laminated with poly(lactic-co-glycolic acid) (PLGA). The PPy-PCL has a bulk conductivity ranging 10-20 S/cm and loses 40 wt % after 7 months under physiologic conditions. Dorsal root ganglia (DRG) grown on flat PPy-PCL/PLGA material exposed to direct current electric fields (EF) of 100 mV/cm for 2 h increased axon growth by 13% (± 2%) toward either electrode of a 2-electrode setup, compared with control grown on identical substrates without EF exposure. Alternating current increased axon growth by 21% (±3%) without an observable directional preference, compared with the same control group. The results from this study demonstrate PLGA-coated PPy-PCL is a unique biodegradable material that can deliver substrate EF stimulation to improve axon growth for peripheral nerve repair.