Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Time dependent accuracy of dental radiometers
Danijela Marović1, Sanja Matić, Katarina Kelić
1Department of Endodontics and Restorative Dentistry, School of Dental Medicine, University of Zagreb, Croatia. marovic@sfzg.hr
Acta Clinica Croatica
|September 24, 2013
Summary
Dental curing light intensity is crucial for restoration quality. Two commercial radiometers (Bluephase meter and Cure Rite) tended to overestimate light output compared to a reference device, with measurement timing impacting accuracy.
Area of Science:
- Dental Materials Science
- Biomedical Optics
- Clinical Dentistry
Background:
- Inadequate polymerization light intensity compromises dental composite restoration quality and longevity.
- Regular control of photopolymerization units is essential for maintaining optimal dental device strength.
Purpose of the Study:
- To compare the accuracy of two handheld radiometers (Bluephase meter and Cure Rite) against a reference integrating sphere.
- To evaluate the influence of measurement time points (initial, 10s, 40s) on light intensity readings from dental curing lights.
Main Methods:
- Light intensity measurements were taken from 16 halogen and 8 LED curing lights.
- Measurements were performed using two commercial radiometers (Bluephase meter, Cure Rite) and an integrating sphere (reference).
- Readings were recorded at the start, 10 seconds, and 40 seconds after illumination initiation.
Main Results:
- Both commercial radiometers (Bluephase meter, Cure Rite) generally overestimated light intensity compared to the integrating sphere.
- The Bluephase meter showed significantly higher values initially and at 10 seconds versus the reference.
- The Cure Rite showed higher values at 10 and 40 seconds compared to the reference; measurement time influenced readings.
Conclusions:
- Commercial radiometers may inaccurately assess dental curing light intensity, potentially overestimating output.
- The time point of measurement significantly affects the recorded light intensity values.
- Radiometer heating is a potential cause for observed inaccuracies in light intensity measurements.
Related Concept Videos
Electronic Distance Measuring Instruments
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over short distances...
Radioactive Decay and Radiometric Dating
Radioactivity is a spontaneous disintegration of an unstable nuclide and is a random process, as all the nuclei in the sample do not decay simultaneously. The number of disintegrations per unit time is called the activity (A), which is directly proportional to the number of nuclei in the sample. The decay constant (λ) is an average probability of decay per nucleus in unit time.