A Model of Long-Term Ventricular Fibrillation in Isolated Rat Hearts
Published on: February 17, 2023
Low-tilt monophasic and biphasic waveforms compared with standard biphasic waveforms in the transvenous
Johan R Bennett1, Karen M Darragh, Simon J Walsh
1Heart Centre, Royal Victoria Hospital, Belfast, UK; Queen's University, Belfast, UK.
Background:
Commercially available implantable defibrillators utilize a high-tilt waveform. Studies in atrial fibrillation and transthoracic defibrillation of ventricular fibrillation (VF) have shown improved defibrillation efficacy using low-tilt (LT) waveforms. We investigated the feasibility, efficacy, and safety of a LT waveform in the transvenous defibrillation of VF and hypothesized that it would be more efficacious than standard tilted biphasic (STB) waveforms.
Methods And Results:
The investigation was performed in four phases in a porcine model: an efficacy study of LT monophasic waveforms (n = 9), an efficacy study of LT biphasic waveforms (n = 9), a comparison study between the most successful LT waveforms and clinically available STB waveforms (n = 15), and a safety study (n = 9). A total of 1,056 shocks were delivered (phase 1: 288, phase 2: 288, phase 3: 480). The LT biphasic 8/4-ms waveform was significantly more likely to successfully defibrillate than the LT monophasic and STB waveforms with an odds ratio of 122.3 (95% confidence interval: 32.5, 460.2, P < 0.001). The calculated defibrillation threshold (E50) for the LT 8/4-ms waveform was 12.7 J compared to 43.5 J and 45.5 J for STB waveforms 6/6 ms and 8/4 ms, respectively, and 47.7 J for LT 12-ms waveform. The LT 8/4-ms waveform had no lasting detrimental effect on cardiac function, and any transient hemodynamical or biochemical changes observed were comparable to those observed with STB waveforms.
Conclusion:
LT waveforms are effective and appear safe in transvenous defibrillation in a porcine model of VF. The LT biphasic 8/4-ms waveform is more efficacious than conventional waveforms.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Voltammetric Techniques: Pulse Voltammetry
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Dysrhythmias VI: Management of Dysrhythmias