Hb H Interference on Measurement Of HbA1c With Ion-Exchange HPLC
Mehmet Agilli1, Halil Yaman, Fevzi Nuri Aydinl
1Department of Medical Biochemistry, Gulhane Military Medical Academy, Ankara, Turkey.
Abstract:
In this article, an interference caused by hemoglobin H (Hb H), during the measurement of hemoglobin A1c (HbA1c) with ion exchange high pressure liquid chromatography (HPLC) method, was presented in blood sample of a 20-year-old male patient. HbA1c measurement was performed with Agilent 1200 HPLC system using a commercial Recipe HbA1c ion-exchange column. Hemoglobin electrophoresis was performed with Interlab G26 agarose electrophoresis automated compact system. HbA1c level was 18.2% and HbA0 level was 81.5% with ion-exchange HPLC method. Patient's fasting serum glucose was assessed before HbA1c measurement and the result was 165 mg/dL (9.16 mmol/L). On the other hand, the result of HbA0 was 87.9%, Hb H was 10.8% and Hb A2 was 1.3% with electrophoresis. Whole blood test values were within reference ranges except MCV. MCV value was 79.6 fL. It is important to keep in mind that HbA1c level might be considered falsely high with ion-exchange HPLC method because of Hb H containing sample.
Related Concept Videos
Ion-Exchange Chromatography
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors