Related Experiment Video
Updated: May 6, 2026

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Intraoperative refractive biometry for predicting intraocular lens power calculation after prior myopic refractive
Tsontcho Ianchulev1, Kenneth J Hoffer2, Sonia H Yoo3
1University of California at San Francisco, San Francisco, California.
Purpose:
To evaluate a new method of intraoperative refractive biometry (IRB) for intraocular lens (IOL) power calculation in eyes undergoing cataract surgery after prior myopic LASIK or photorefractive keratectomy.
Design:
Retrospective consecutive cases series.
Participants:
We included 215 patients undergoing cataract surgery with a history of myopic LASIK or photorefractive keratectomy.
Methods:
Patients underwent IRB for IOL power estimation. The Optiwave Refractive Analysis (ORA) System wavefront aberrometer was used to obtain aphakic refractive measurements intraoperatively and then calculate the IOL power with a modified vergence formula obtained before refractive surgery. Comparative effectiveness analysis was done for IRB predictive accuracy of IOL power determination against 3 conventional clinical practice methods: surgeon best preoperative choice (determined by the surgeon using all available clinical data), the Haigis L, and the Shammas IOL formulas.
Main Outcome Measures:
Median absolute error of prediction and percentage of eyes within ±0.50 diopters (D) and ±1.00 D of refractive prediction error.
Results:
In 246 eyes (215 first eyes and 31 second eyes) IRB using ORA achieved the greatest predictive accuracy (P < 0.0001), with a median absolute error of 0.35 D and mean absolute error of 0.42 D. Sixty-seven percent of eyes were within ±0.5 D and 94% were within ±1.0 D of the IRB's predicted outcome. This was significantly more accurate than the other preoperative methods: Median absolute error was 0.6, 0.53, and 0.51 D for surgeon best choice, Haigis L method, and Shammas method, respectively.
Conclusions:
The IOL power estimation in challenging eyes with prior LASIK/photorefractive keratectomy was most accurately predicted by IRB/ORA.