Effect of dynamic random leaks on the monitoring accuracy of home mechanical ventilators: a bench study
Ana Sogo, Jaume Montanyà, Eduard Monsó
1Department of Pneumology, Corporació Sanitària Parc Taulí, Institut Universitari Parc Taulí, Universitat Autònoma de Barcelona, Sabadell, Spain. mlujan@tauli.cat.
Background:
So far, the accuracy of tidal volume (VT) and leak measures provided by the built-in software of commercial home ventilators has only been tested using bench linear models with fixed calibrated and continuous leaks. The objective was to assess the reliability of the estimation of tidal volume (VT) and unintentional leaks in a single tubing bench model which introduces random dynamic leaks during inspiratory or expiratory phases.
Methods:
The built-in software of four commercial home ventilators and a fifth ventilator-independent ad hoc designed external software tool were tested with two levels of leaks and two different models with excess leaks (inspiration or expiration). The external software analyzed separately the inspiratory and expiratory unintentional leaks.
Results:
In basal condition, all ventilators but one underestimated tidal volume with values ranging between -1.5 ± 3.3% to -8.7% ± 3.27%. In the model with excess of inspiratory leaks, VT was overestimated by all four commercial software tools, with values ranging from 18.27 ± 7.05% to 35.92 ± 17.7%, whereas the ventilator independent-software gave a smaller difference (3.03 ± 2.6%). Leaks were underestimated by two applications with values of -11.47 ± 6.32 and -5.9 ± 0.52 L/min. With expiratory leaks, VT was overestimated by the software of one ventilator and the ventilator-independent software and significantly underestimated by the other three, with deviations ranging from +10.94 ± 7.1 to -48 ± 23.08%. The four commercial tools tested overestimated unintentional leaks, with values between 2.19 ± 0.85 to 3.08 ± 0.43 L/min.
Conclusions:
In a bench model, the presence of unintentional random leaks may be a source of error in the measurement of VT and leaks provided by the software of home ventilators. Analyzing leaks during inspiration and expiration separately may reduce this source of error.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...