Pacemaker clinic evaluations: key to early identification of surgical problems
Abstract:
The pacemaker center evaluation was responsible for the timely reoperation of 341 pacemaker patients over the last three years. The most common indication for reoperation was battery end-of-service (46.3%). Battery testing and maintenance of accurate records for trend analysis ensures prompt generator replacement. Atrial and/or ventricular lead malfunction was the second largest indication (26.3%). Lead malfunction detected by bracketing capture and sensing thresholds included: dislodgement, penetration, exit block, fracture, insulation failure, and abnormal sensing. The third largest indication for reoperation was pacemaker pocket erosion and/or infection (15.5%). Examination of the pacemaker pocket site is an integral part of an evaluation. In conclusion, the three largest indications for reoperations can be best demonstrated in a pacemaker center where evaluations are personal and thorough, and accurate record keeping is maintained.