Related Experiment Video
Updated: May 4, 2026

Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Resistance to secondary thrombosis of the On-X mitral prosthesis
Federica Jiritano1, Giuseppe F Serraino2, Michele Rossi2
1Cardiac Surgery Unit, Magna Graecia University of Catanzaro, Italy. fede.j@hotmail.it
Abstract:
A 60-year-old man underwent aortic and mitral valve replacement with mechanical On-X prostheses, plus tricuspid annuloplasty ring and maze radiofrequency ablation. Two years later, he was admitted to hospital with severe orthopnea and hemolysis. Echocardiography revealed a blocked prosthetic mitral leaflet due to interference of the preserved subvalvular apparatus. Intraoperatively, one leaflet was found to be blocked by the native mitral apparatus, but no thrombus and/or pannus was present. Although, despite an abnormal blood flow, the On-X prosthesis is resistant to secondary thrombosis, even under adverse hemodynamic conditions, interference of the valve with cardiac structures remains a possibility.