Related Experiment Video
Updated: May 4, 2026

An All-on-chip Method for Rapid Neutrophil Chemotaxis Analysis Directly from a Drop of Blood
Published on: June 23, 2017
A capillary dielectrophoretic chip for real-time blood cell separation from a drop of whole blood
Shu-Hsien Liao1, Ching-Yu Chang2, Hsien-Chang Chang3
1Department of Biomedical Engineering, National Cheng Kung University, Tainan, Taiwan.
Abstract:
This study proposes a capillary dielectrophoretic chip to separate blood cells from a drop of whole blood (approximately 1 μl) sample using negative dielectrophoretic force. The separating efficiency was evaluated by analyzing the image before and after dielectrophoretic force manipulation. Blood samples with various hematocrits (10%-60%) were tested with varied separating voltages and chip designs. In this study, a chip with 50 μm gap design achieved a separation efficiency of approximately 90% within 30 s when the hematocrit was in the range of 10%-50%. Furthermore, glucose concentration was electrochemically measured by separating electrodes following manipulation. The current response increased significantly (8.8-fold) after blood cell separation, which was attributed not only to the blood cell separation but also to sample disturbance by the dielectrophoretic force.
More Related Videos
09:45Microfluidics-based High-throughput Circulating Tumor Cell Sorting and Single-cell Sequencing Technology
Published on: November 14, 2025
08:33Microfluidic Device for the Separation of Non-Metastatic MCF-7 and Non-Tumor MCF-10A Breast Cancer Cells Using AC Dielectrophoresis
Published on: August 11, 2022
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...