Comparison of automated measurements of electrocardiographic intervals and durations by computer-based algorithms of
Paul Kligfield1, Fabio Badilini2, Ian Rowlandson3
1Division of Cardiology, Weill Cornell Medical College, New York, NY.
Background And Purpose:
Automated measurements of electrocardiographic (ECG) intervals are widely used by clinicians for individual patient diagnosis and by investigators in population studies. We examined whether clinically significant systematic differences exist in ECG intervals measured by current generation digital electrocardiographs from different manufacturers and whether differences, if present, are dependent on the degree of abnormality of the selected ECGs.
Methods:
Measurements of RR interval, PR interval, QRS duration, and QT interval were made blindly by 4 major manufacturers of digital electrocardiographs used in the United States from 600 XML files of ECG tracings stored in the US FDA ECG warehouse and released for the purpose of this study by the Cardiac Safety Research Consortium. Included were 3 groups based on expected QT interval and degree of repolarization abnormality, comprising 200 ECGs each from (1) placebo or baseline study period in normal subjects during thorough QT studies, (2) peak moxifloxacin effect in otherwise normal subjects during thorough QT studies, and (3) patients with genotyped variants of congenital long QT syndrome (LQTS).
Results:
Differences of means between manufacturers were generally small in the normal and moxifloxacin subjects, but in the LQTS patients, differences of means ranged from 2.0 to 14.0 ms for QRS duration and from 0.8 to 18.1 ms for the QT interval. Mean absolute differences between algorithms were similar for QRS duration and QT intervals in the normal and in the moxifloxacin subjects (mean ≤6 ms) but were significantly larger in patients with LQTS.
Conclusions:
Small but statistically significant group differences in mean interval and duration measurements and means of individual absolute differences exist among automated algorithms of widely used, current generation digital electrocardiographs. Measurement differences, including QRS duration and the QT interval, are greatest for the most abnormal ECGs.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....