Evaluation of the difference-correction effect of the gamma camera systems used by easy Z-score Imaging System (eZIS)
Yasushi Yamamoto1, Masahisa Onoguchi, Kazunori Kawakami
1Department of Radiology, Shimane University Hospital, 89-1 Enya-cho, Izumo, Shimane, 693-8501, Japan, yasushi@med.shimane-u.ac.jp.
Objective:
We examined the difference of the effect by data to revise a gamma camera difference. The difference-correction method of the camera is incorporated in eZIS analysis.
Methods:
We acquired single photon emission computed tomography (SPECT) data from the three-dimensional (3D) Hoffman brain phantom (Hoffman), the three-dimensional brain phantom (3D-Brain), Pool phantom (pool) and from normal subjects (Normal-SPECT) to investigate compensating for a difference in gamma camera systems. We compared SPECT counts of standard camera with the SPECT counts that revised the difference of the gamma camera system (camera). Furthermore, we compared the "Z-score map (Z-score)". To verify the effect of the compensation, we examined digitally simulated data designed to represent a patient with Alzheimer's dementia. We carried out both eZIS analysis and "Specific Volume of interest Analysis (SVA)".
Results:
There was no great difference between the correction effect using Hoffman phantom data and that using 3D-Brain phantom data. Furthermore, a good compensation effect was obtained only over a limited area. The compensation based on the pool was found to be less satisfactory than any of the other compensations according to all results of the measurements examined in the study. The compensation based on the Normal-SPECT data resulted in a Z-score map (Z-score) for the result that approximated that from the standard camera. Therefore, we concluded that the effect of the compensation based on Normal-SPECT data was the best of the four methods tested.
Conclusions:
Based on eZIS analysis, the compensation using the pool data was inferior to the compensations using the other methods tested. Based on the results of the SAV analysis, the effect of the compensation using the Hoffman data was better than the effect of the compensation using the 3D-Brain data. By all end-point measures, the compensation based on the Normal-SPECT data was more accurate than the compensation based on any of the other three phantoms.
More Related Videos
10:23Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Introduction to z Scores
z scores...
Introduction to z Scores
z scores...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...