Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
Fully automated on-line two-dimensional liquid chromatography in combination with ESI MS/MS detection for
Kristaps Klavins1, Dinh Binh Chu, Stephan Hann
1University of Natural Resources and Life Sciences, BOKU-Vienna, Department of Chemistry, Division of Analytical Chemistry, Muthgasse 18, 1190, Vienna, Austria. gunda.koellensperger@boku.ac.at.
Abstract:
A mass spectrometric quantitative assay was developed for the analysis of 10 sugar phosphates in the yeast Pichia pastoris. As a novelty, two-dimensional chromatography based on a fully automated heart-cutting LC-LC technique was introduced. Using a ten-port valve, ten fractions of the first chromatographic dimension, i.e. anion exchange chromatography (AEC), were transferred and separated by the orthogonal second dimension, i.e. separation on porous graphitized carbon. The chromatographic separation on the second dimension was optimized for each transferred fraction minimizing the separation time and ensuring complete removal of the salt constituents of the AEC eluents. The latter being crucial for electrospray mass spectrometric detection was confirmed by combining the LC-LC separation with on-line ICP-MS detection. These measurements showed that sodium elution was completed after 0.8 min. Consequently, an analysis time of 1 min per transferred peak was established. In this way, the excellent peak capacity given by ion exchange could be conserved in the second dimension at the same time enabling mass spectrometric detection. Sub-μM limits of detection could be obtained by the new LC-LC-MS/MS methods ranging between 0.03 and 0.19 μM for the investigated compounds (only 3GAP showed a LOD of 1 μM). The method was applied to the quantification of ten sugar phosphates in yeast extracts utilizing internal standardization with a fully labeled (13)C yeast extract. Typically, the standard uncertainties for N = 3 replicates assessed by the LC-LC-MS/MS set-up were <5%.
More Related Videos
07:25Quantitative Metabolomics of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: January 5, 2021
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021