Control point analysis comparison for 3 different treatment planning and delivery complexity levels using a
Ady Abdellatif1, Stewart Gaede2
1Department of Physics and Engineering, London Regional Cancer Program, London, Ontario, Canada.
Summary
Control Point Analysis effectively evaluates volumetric-modulated arc therapy (VMAT) plan delivery accuracy. Prostate VMAT plans showed higher accuracy, while SBRT and head-and-neck plans indicated reduced accuracy per control point.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Quality Assurance
Background:
- Volumetric-modulated arc therapy (VMAT) is a complex radiotherapy technique.
- Accurate dose delivery verification is crucial for patient safety and treatment efficacy.
- Control Point Analysis offers a method to assess VMAT delivery accuracy at a granular level.
Purpose of the Study:
- To evaluate the efficacy of Control Point Analysis for assessing VMAT plan delivery.
- To compare the delivery accuracy of VMAT plans across different treatment sites (lung SBRT, head-and-neck, prostate) and complexity levels.
- To investigate the impact of control point binning on gamma analysis results.
Main Methods:
- Thirty patient VMAT plans (10 lung SBRT, 10 head-and-neck, 10 prostate) were analyzed.
- Plans were delivered on a Varian linear accelerator and measured using an ArcCHECK phantom.
- Delivered dose was compared to planned dose using Sun Nuclear Patient 6 software with Control Point Analysis and gamma passing criteria (e.g., 3%, 3mm and 2%, 2mm).
Main Results:
- Prostate VMAT plans demonstrated the highest gamma passing percentages (e.g., 95.4%–98.3% for 3%, 3mm).
- Increasing the number of binned control points improved passing rates, particularly for the 2%, 2mm criteria.
- Lung SBRT (high MU/control point) and head-and-neck (high modulation) plans showed lower delivery accuracy per control point compared to prostate plans.
Conclusions:
- Control Point Analysis is a valuable tool for VMAT plan quality assurance.
- Treatment complexity, characterized by Monitor Unit (MU)/control point or modulation degree, impacts delivery accuracy.
- Prostate VMAT plans exhibit superior delivery accuracy compared to lung SBRT and head-and-neck VMAT plans.