Simulation of response functions of fast neutron sensors and development of thin neutron silicon sensor
Masashi Takada1, Takashi Nakamura2, Mikihiko Matsuda3
1National Institute of Radiological Sciences, Chiba 263-8555, Japan m_takada@nirs.go.jp.
Abstract:
On radiation detection using silicon sensor, signals are produced from collected charges in a depletion layer; however, for high-energy particles, this depletion layer is extended due to funnelling phenomenon. The lengths of charge collection were experimentally obtained from proton peak energies in measured pulse-heights. The length is extended with increasing proton energy of up to 6 MeV, and then, is constant over 6 MeV. The response functions of fast neutron sensors were simulated for 5- and 15-MeV monoenergetic and (252)Cf neutron sources using the Monte Carlo N-Particle eXtended code. The simulation results agree well with the experimental ones, including the effect of funnelling phenomenon. In addition, a thin silicon sensor was developed for a new real-time personal neutron dosemeter. Photon sensitivity is vanishingly smaller than neutron one by a factor of 5×10(-4).