Highly Sensitive and Rapid Fluorescence Detection with a Portable FRET Analyzer
Published on: October 1, 2016
Portable system for the detection of micromolar concentrations of glucose
Yordan Kostov1, Xudong Ge1, Govind Rao1
1Center for Advanced Sensor Technology, Department of Chemical, Biochemical and Environmental Engineering, University of Maryland Baltimore County, 1000 Hilltop Circle, Baltimore MD, 21250.
Abstract:
Glucose in non-invasively collected biofluids is generally in the micromolar range and thus, requires sensing methodologies capable of measuring glucose at these levels. Here, we present a small fluorometer system that can quantify glucose in the range of 0-5 μM with resolution of ~0.07 μM. It relies on the glucose binding protein (GBP) fluorescently labeled with two fluorophores. Fluorescence signals from the dual-labeled GBP are utilized in a ratiometric mode, making the measurements insensitive to variations in protein concentration and other systematic errors. Fluorescence is quantified by a miniature, dedicated ratiometric fluorometer that is powered via USB. Concentration is calculated using an ultra-mobile personal computer (UMPC). The whole system is designed to be pocket sized suitable for point-of-care or bedside applications. Test results suggest that the system is a promising tool for accurate measurements of low glucose concentrations (0.1-10 μM) in biological samples.