Related Experiment Video
Updated: May 2, 2026

07:56
An Automated System for Sound Localization Testing in Hearing-Impaired Listeners
Published on: March 13, 2026
184
An assistive device for direction estimation of a sound source
Assistive Technology : the Official Journal of RESNA
|March 14, 2014
Summary
This study introduces eyeglasses that visually guide hearing-impaired individuals toward sound sources. The sound pressure level comparison method achieved 92% accuracy in directional estimation.
Area of Science:
- Assistive Technology
- Biomedical Engineering
- Acoustics
Background:
- Hearing impairment affects sound localization.
- Existing assistive devices may not be discreet.
- Visual cues can aid in sound source identification.
Purpose of the Study:
- To develop and evaluate a wearable device for visual sound source indication.
- To provide discreet assistance for individuals with hearing loss.
- To compare directional estimation methods for accuracy.
Main Methods:
- Acoustical data acquired via microelectromechanical-system microphone array on eyeglasses.
- Real-time sound source direction estimation using two methods: delay-and-sum beamformer and sound pressure level (SPL) comparison.
- Performance evaluated using a head-and-torso simulator.
Main Results:
- The sound pressure level (SPL) comparison method demonstrated 92% accuracy.
- The delay-and-sum beamformer method achieved 84% accuracy.
- The device is designed as common eyeglasses to minimize attention to hearing impairment.
Conclusions:
- The developed eyeglasses effectively provide visual cues for sound source direction.
- The SPL comparison method offers superior accuracy for this application.
- This discreet wearable device shows promise for enhancing environmental awareness in the hearing impaired.
Related Concept Videos
Echo
1.2K
The human ear cannot distinguish between two sources of sound if they happen to reach within a specific time interval, typically 0.1 seconds apart. More than this, and they are perceived as separate sources.
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
1.2K
Electronic Distance Measuring Instruments
771
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over...
771
Auditory Perception
1.5K
The auditory system is essential for sound perception, utilizing various critical structures. When sound waves enter the outer ear, they travel through the ear canal and cause the eardrum to vibrate. These vibrations are then transmitted to the middle ear, where three tiny bones – the malleus, incus, and stapes – amplify the sound. This amplification is crucial, as it ensures that the sound vibrations are strong enough to be conveyed to the inner ear. These vibrations then reach the...
1.5K
Equipments Used To Measure Blood Pressure
3.4K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K

