Related Experiment Videos
Impedance changes during catheter ablation and their relationship to electrical arcing and clinical efficacy
D Cunningham1, A J Ahsan, E Rowland
1National Heart Hospital, London, UK.
Abstract:
This study has used high energy ablative shocks in saline and in man to characterize the complex, energy dependent behavior of the impedance at the electrode-tissue interface. In vitro shocks showed a fall in impedance when an arc was formed, due to formation around the electrode of a vapor globe rendered conductive by the arc. Following arc extinction, this same vapor globe insulates the electrode, and subsequently collapses, forming a shock wave, and allowing current flow to resume. Shocks delivered to tissue specimens exhibited progressively higher impedance and less insulation as contact pressure was increased. In vivo, lesser degrees of insulation during arcing shocks were associated with greater clinical efficacy.